Type of Product:Device
Medsafe Reference:37614
Brand Name:Aisys Anaesthesia Work Stations
Model:Aisys CS2, Aisys CS2 with Et Control and S/5 Aisys Anesthesia Delivery Systems
Affected:Replaceable parts: 1011-8000-000-S - Gas Mixer Electronic – Assembly. 5700665 - R-FMI34141 GAS MIXER ELECTRONIC
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware that Aisys CS2, Aisys CS2 with Et Control and S/5 Aisys Anesthesia Delivery Systems, containing certain electronic gas mixer assemblies, can automatically transition to Alternate O2 control.
In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms ("Set Alt O2 Flow! Agent delivery off!" or "Set Alt O2 flow! Check agent setting!").
When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/9/2026