Detail of Recall

Type of Product:Device
Medsafe Reference:37605
Brand Name:Aspire Alternating Air Mattress
Model:Aidacare Aspire Active Air 8 Pressure Care Air Mattress
Affected:Item Numbers: PTM639300 & PTM639310 - All serial numbers
Software version:
Recalling Organisation:Compliance Management Solutions NZ
13 Fernbank Lane
Greenhithe
North Shore City
Auckland 0632
Contact Information:
Manufacturer:Aidacare Pty Ltd
Issue:The manufacturer has updated the user manual for these mattresses. The device remains safe to use, has not changed, and no return, repair or modification is required. The updated manual provides enhanced guidance to support the continued safe and effective use of the product.
The key updates include:
- Additional warnings when using the device in the presence of oxygen enriched environments, including oxygen concentrators.
- Additional setup instructions, in particular correct air-supply hose routing to not lift the mattress off the bed base, and when using bolsters to extend bed lengths.
Improper set up and use of the device may result in electrostatic discharge, which may potentially cause a fire in oxygen-enriched environments.
Recall action type:Alert
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/9/2026