Detail of Recall

Type of Product:Device
Medsafe Reference:37602
Brand Name: LABScreen™ Mixed Class I & II
Model:LABScreen™ Mixed Class I & II (Cat ID. LSM12, Lot 026)
Affected:Multiple lots
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:One Lambda Inc
Issue:The Data Sheets and Catalog Files associated with LABScreen Mixed Class I & II (Cat ID. LSM12, Lot 026) have the incorrect Cell ID and panel typing information for Bead ID 17. The correct information for Bead 17 is Cell ID C4966 (Panel Typing A2, B46, Bw6, Cw1) rather than Cell ID G0223 (Panel Typing A1, A11, B46, B57, Bw6, Bw4, Cw6, Cw8). There is no risk to patient or end user. The performance of the product to qualitatively detect HLA Class I antibody reactivity is not affected.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/9/2026