Detail of Recall

Type of Product:Device
Medsafe Reference:37599
Brand Name:Olympus - Ligating device, single-use (Japan)
Model:SINGLE USE LIGATING DEVICE POLYLOOP (HX-400U-30)
Affected:Multiple Lot numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer previously communicated a Product Correction (Medsafe Ref #36740) related to additional clarifications in the Instructions for Use (IFU) for proper release of the Single Use Ligating Device (HX-400U-30). Alternatively, customers may have received a Notification of IFU Change with recently purchased products. The action was taken based on complaints from customers indicating that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally anchored in place around patient anatomy. The manufacturer is introducing a design change to address complaints that the ligation loop was unable to release or detach as expected during use, with operation pipe breakage being one of the contributing factors to these complaints. The manufacturer revised the method used to secure the guide pipe. As a result, even if the sheath near the handle is not fully straightened, the operation pipe can be prevented from bending.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/9/2026