Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37590
Brand Name:Shower Trolley: Carevo
Model:
Affected:All Carevo devices manufactured between 2015-Aug-01 and 2024-Feb-20
Software version:
Recalling Organisation:Arjo New Zealand Limited
34 Vestey Drive
Mt Wellington
AUCKLAND 1060
Contact Information:0800 000 151
Manufacturer:ArjoHuntleigh AB
Issue:Unintentional side support opening does not occur spontaneously. Internal investigations found that the primary risk arises when the handle is not fully locked and additional force is applied during patient handling, increasing the potential for a patient fall. To address this, the Instructions for Use (IFU) have been updated with additional warnings on proper handle engagement and ergonomic handling practices. The investigation also identified that a specific supplier batch of torsion springs may not consistently return the handle to its fully locked position, increasing the likelihood of incomplete handle engagement and, consequently, unintentional side support opening under certain conditions.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/9/2026