Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37552 |
| Brand Name: | SSY,USM,IS4000, da Vinci Xi Surgical System, da Vinci X Surgical System |
| Model: | ASSY,USM,IS4000, da Vinci Xi Surgical System, da Vinci X Surgical System |
| Affected: | Product Codes: 380647 |
| Software version: | |
| Recalling Organisation: | Device Technologies New Zealand 47 Arrenway Drive Albany AUCKLAND 0632 |
| Contact Information: | 09 913 2000 |
| Manufacturer: | Intuitive Surgical Inc |
| Issue: | The manufacturer has identified that the screws on the cannula mount of the USM are from a specific population that are susceptible to breaking over time. When the cannula mount experiences a breakage of all six screws, the cannula mount can separate from the rest of the USM. Breakage of any fewer than six screws will not result in cannula mount separation, or impact to clinical performance. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/9/2026 |