Type of Product:Device
Medsafe Reference:37545
Brand Name:Physio-Control LIFEPAK 35 defibrillator/monitor
Model:LIFEPAK 35 Monitor/Defib
Affected:Catalogue number - 70335-000016, 70335-000017, 70335-000030, 70335-000018, 70335-000034
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Physio-Control Corporation
Issue:The manufacturer has identified a problem with the LP35 where the device does not display the message "CHECK RHYTHM / ECG Analysis Inconclusive" as intended when all below conditions are present simultaneously. The problem does not affect the display of "Shock Advised" or "No Shock Advised" messages, and therapy delivery remains available. Additionally, this nonconformance is limited to the device on-screen prompts only and rhythm analysis is not impacted by this problem.
Problem 1: Under a specific combination of non-default Manual Mode settings, the "CHECK RHYTHM / ECG Analysis Inconclusive" prompt may not be displayed. Device functionality, rhythm analysis, and therapy delivery are unaffected. Problem
2: Certain printer and modem access port connectors may not have been fully seated during manufacturing, resulting in improper connector positioning within the rear housing assembly.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/9/2026