Detail of Recall

Type of Product:Device
Medsafe Reference:37495
Brand Name:Intelerad InteleradPACS - InteleConnect Portal / TechPortal Version
Model:IntelePACS – InteleConnect Portal / Tech Portal Version
Affected:
Software version:
Recalling Organisation:Clinical & Regulatory Services Limited
30 Fulford Road
Havelock North
Hastings 4130
Contact Information:(06) 844 4490
Manufacturer:Intelerad Medical Systems Incorporated
Issue:The manufacturer has become aware of a problem affecting certain versions of IntelePACS -InteleConnect Portal and Technologist Portal. Following an update to Chromium-based web browsers (for example, Google Chrome 152, Microsoft Edge 152) in late August, the Physician Portal Case Viewer may display a report from a previously viewed study alongside images from the currently selected study. The report panel may not refresh correctly when navigating between studies. Both the report and the study may appear under the same patient identifiers without any indication that the report does not match the displayed images. This could result in misdiagnosis, delayed or incorrect patient treatment.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/9/2026