Type of Product:Device
Medsafe Reference:37475
Brand Name:Medifab Safe Surround Genie Bed
Model:Safe Surround Genie Bed
Affected:Product Code: 7959-0016-000 Manufactured on or before 9 July 2025
Software version:
Recalling Organisation:Medifab Limited
1 Pereita Drive
Rolleston 7675
Contact Information:0800 543 343
Manufacturer:KayserBetten GmbH & Co KG
Issue:The manufacturer has identified that the hand control may not remain locked after use. If the hand control is not locked while a user is unattended, unintended movement of the bed may occur, which could result in entrapment and serious injury or death.
Two deaths have been reported overseas in relation to this problem.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Public Recall
Date Commenced:9/10/2026