Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37475 |
| Brand Name: | Medifab Safe Surround Genie Bed |
| Model: | Safe Surround Genie Bed |
| Affected: | Product Code: 7959-0016-000 Manufactured on or before 9 July 2025 |
| Software version: | |
| Recalling Organisation: | Medifab Limited 1 Pereita Drive Rolleston 7675 |
| Contact Information: | 0800 543 343 |
| Manufacturer: | KayserBetten GmbH & Co KG |
| Issue: | The manufacturer has identified that the hand control may not remain locked after use. If the hand control is not locked while a user is unattended, unintended movement of the bed may occur, which could result in entrapment and serious injury or death. Two deaths have been reported overseas in relation to this problem. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Public Recall |
| Date Commenced: | 9/10/2026 |