| Type of Product: | Device |
| Medsafe Reference: | 37474 |
| Brand Name: | Carestation 850 and 850c Anesthesia Systems |
| Model: | Carestation 850 and 850c Anesthesia Systems: Ref #1012-9850-000 and 1012-9855-000 |
| Affected: | All Carestation 850/850c Anesthesia Delivery Systems with software versions 01SP05, 01SP06, 01SP07, 01SP08. |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | The manufacturer has become aware of the potential unexpected shutdown of Carestation 800 Series Anesthesia Systems, if the AC mains power is unplugged or AC mains power fails. This can occur following an incomplete system startup occurring when the ON/Standby switch is not pushed long enough as indicated in the User Reference Manual. Due to this issue, if AC mains power is interrupted for any reason, the system could shut down disrupting gas flow, mechanical ventilation, manual ventilation, and anesthetic agent delivery. If this is not promptly addressed by the user, the loss of ventilation can be life threatening. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/9/2026 |