Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37451
Brand Name:DxC 500i Clinical Analyzer
Model:With ISE DxC 500 AU Module w/ISE, DxC 500i and Without ISE DxC 500 AU Module, DxC 500i
Affected:Software - 1.3.5 or earlier
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has identified a software issue affecting DxC 500i Clinical Analyzer when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyser to revert the test units to the system default setting without notifying the user.
The system does not provide a warning or notification to alert the user that the configured units have reverted to the default units.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/9/2026