Database updated: 20 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37400
Brand Name:NOXBOXi Nitric Oxide Monitor and Delivery System
Model:NOxBOXi Intelligent Delivery and Monitor
Affected:Product code: NOXBOX-1
Software version:
Recalling Organisation:Keyport Ltd
431 Te Rapa Road
HAMILTON
Contact Information:
Manufacturer:NOxBOX Ltd
Issue:The manufacturer has identified an increased trend in complaints for oxygen leaks within the NOxMixer module of the NOxBOXi device due to failure of the tubing or tubing connection. The NOxMixer module is intended to deliver a continuous flow of nitric oxide (NO), mixed in line with oxygen (O2) for emergency applications and when manually ventilating a patient (i.e., during Emergency and Manual Bagging modes). This failure occurs inside the NOxMixer module when a leak develops in the internal oxygen tubing or the tubing connection at the elbow fitting upstream of the oxygen flowmeter.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/8/2026
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /