Detail of Recall

Type of Product:Device
Medsafe Reference:37394
Brand Name:Dexcom CGM system
Model:SW12300 G7 App (iOS) – Version 2.12
Affected:In G7 iOS App version 2.12, only Critical Alerts are designed to override the phone's Silent Mode or Focus Mode setting.
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Dexcom Inc
Issue:In G7 iOS App version 2.12, only Critical Alerts are designed to override the phone's Silent Mode or Focus Mode setting. Noncritical alerts behave in accordance with the phone's notification settings. In the affected version of the Dexcom G7 iOS app, non-critical alerts may not sound if the phone is in Silent Mode or Focus Mode. If a user is unaware that non-critical alerts may not sound due to behaving in accordance with the phone settings, there is a potential for delayed detection of a hypoglycaemic or hyperglycaemic event.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/8/2026