Detail of Recall

Type of Product:Device
Medsafe Reference:37393
Brand Name:Dexcom CGM system
Model:SW12299 G7 App (Android) – Version 2.9.0 - 2.11.2 and SW12300 G7 App (iOS) – Version 2.9.0 - 2.11.2
Affected:Dexcom G7 App (iOS and Android), Versions 2.9-2.11.2, and the suppression of critical alerts.
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Dexcom Inc
Issue:The manufacturer has identied that when in 'Do Not Disturb' mode, critical alerts may not vibrate as expected and 'disk space critical,' 'recoverable error,' and 'Unrecoverable error' alerts may not sound or vibrate for up to 25 minutes. You're using an older version of the Dexcom G7 app, which may not perform as expected due to this issue.
If a user is unaware of a prompt due to an alert not performing as expected, there is a potential for a missed / delayed detection of a hypoglycaemic or hyperglycaemic event.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Retail
Date Commenced:28/8/2026