Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37392 |
| Brand Name: | Brilliance iC, Ingenuity CT/ Ingenuity Core, Ingenuity CT Systems |
| Model: | Brilliance iC, Ingenuity CT/ Ingenuity Core, Ingenuity CT and IQon Spectral CT |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Brand Name: | [MDR] Spectral CT, IQon Spectral CT, Brilliance iCT, Brilliannce iCT Upgrade |
| Model: | Brilliance iCT [iCT Elite with IMR, iCT Elite] Rel 4.1.10 and IQon Spectral CT [IQon Spectral CT Pro, IQon Spectral CT Premium] |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Brand Name: | [MDR] Ingenuity Core, Ingenuity CT, Big Bore, Brilliance CT Big Bore System Upgrade |
| Model: | Ingenuity Core [CT 5000 Ingenuity Plus] Rel 4.1.10, Ingenuity Core 128 [CT 5000 Ingenuity Pro] Rel 4.1.10, Ingenuity CT [CT 5000 Ingenuity Premium] Rel 4.1.10 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | The manufacturer has identified a software issue with the Philips iCT, Ingenuity CT, and Brilliance CT 64 systems that could affect the performance of the equipment. During the cardiac scan planning, if the X (XCA, XCB, XCC, XCD) filter for reconstruction is selected, the reconstructed cardiac images may display unexpected noise. Noisy cardiac images may result in reduced image quality, including recognizable artifacts, anatomy obscured by artifacts, or non-diagnostic images. As an alternative, instead of using X filter for cardiac imaging, you can use CA, CB, CC, and CD filters and evaluate the image quality during live planning or during offline reconstruction. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/8/2026 |