Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37340
Brand Name:Carescape Patient monitor, central unit
Model:Product code: STU CSCS V3 MAS800 Desktop, Product code: STV CSCS V3 MAI800 Integrated
Affected:Part number: 5867474-04, 5867474-013, 5867474-03
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:The manufacturer has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the Allow list database is corrupted message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/8/2026