Type of Product:Device
Medsafe Reference:37339
Brand Name:Clearview hCG Cassette and Innovacon hCG One Step Pregnancy Test Device
Model:Clearview™ hCG Cassette (20 mIU/mL), Innovacon™ hCG One Step Pregnancy Test Device (Urine)
Affected:Part number: 92210, FHC-102 - All lots within expiration date
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Innovacon Inc
Issue:Update 2/9/26: In consideration of the current market shortage of the affected products, this action has been revised for distributors. Distributors may continue to supply inventory to their customers, on the provision that all future customer orders must be accompanied by the Urgent Product Correction Customer Letter and Customer Reply Form.

As part of routine post-market surveillance activities, an investigation was conducted into reports of false-positive results, and determined that while the tests demonstrate expected results at and above the established sensitivity levels, they may detect low levels of hCG below the stated assay sensitivity. This may result in false positive test results in samples from non-pregnant individuals.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional