Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37327 |
| Brand Name: | ImagerTM II Angiographic Catheters |
| Model: | Multiple package codes |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Boston Scientific NZ Ltd 3 Diamond Street Eden Terrace Auckland 1021 |
| Contact Information: | 0800 742 678 |
| Manufacturer: | Boston Scientific Corporation |
| Issue: | The manufacturer has determined that these Catheters were manufactured with reduced levels of stabilizing agents in the tips. This may increase the risk of tip degradation and detachment. Although no increase in tip detachment complaints has been identified for the affected lots, the devices were manufactured outside the approved design specification and are being removed from the market. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/8/2026 |