Detail of Recall

Type of Product:Device
Medsafe Reference:37310
Brand Name:Cardioroot
Model:CARDIOROOT Woven - Model number: HEWROOT0024, HEWROOT0026, HEWROOT0028, HEWROOT0030, HEWROOT0032, HEWROOT0034
Affected:All CARDIOROOT Woven products within labelled shelf life
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Intervascular SAS
Issue:Internal testing of samples from CARDIOROOT vascular grafts produced Integral Water Permeability (IWP) results outside the limit specified in the product's Instructions for Use (IFU) (i.e. <5 ml/cm²/min). Given that the affected products did not meet the specified acceptance criteria for water permeability, they are out of specification and may pose risks to patient health.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/8/2026