Detail of Recall

Type of Product:Device
Medsafe Reference:37309
Brand Name:Automated Impella controller with Impella Connect
Model:Product code: 0042-0040-AU - Optical, AIC, Impella Connect, Pkgd, AU
Affected:All lots and serial numbers
Software version:
Recalling Organisation:Bio-Excel (NZ) Limited trading as Bio-Excel (Australia) Pty Ltd
Unit C6, 9 Tait Place
Albany
Auckland 0632
Contact Information:
Manufacturer:Abiomed Inc
Issue:The manufacturer has identified an increase in complaints due to purge cassette recognition issues resulting from failures of the purge flag component within the purge pressure sensor assembly. The purge flag engages with the purge disc and allows the AIC to recognize that the purge disc is inserted properly. The manufacturer is addressing these failures by replacing the purge flag and purge retainer components within the purge pressure sensor assembly of the AIC.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/8/2026