Detail of Recall

Type of Product:Device
Medsafe Reference:37291
Brand Name:Getinge Assured Batch Monitoring PCD
Model:Model number - PL589/ 6005500575
Affected:54043A
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:SteriTec Products Inc
Issue:This problem was initiated by a customer product complaint regarding a missing O-ring. There have been no reported testing failures or adverse events, and the defect is unlikely to cause harm to the user. The root cause was identified as a combination of a supplier production process failure and an internal quality inspection error.
Failure of the device, due to the missing O-ring, may state sterilisation cycle was not completed and may require re-sterilisation, which may disrupt workflow. This could result in delay of goods and, as such, delay of procedure; should this occur, it is anticipated that any procedural delays would be prior to administration of anaesthesia.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/8/2026