Database updated: 20 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37289
Brand Name:Getinge Steam Sterilizer
Model:HS 6610 AR1, HS 6613 ER-1, HS 6613 AR-2, HS 6613 AC-1, HS 6610 AC-1, HS 6610 EC-1, HS 6610 AR-1, HS 6613 AR-1
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Sterilization AB
Brand Name:LTSF Sterilizer
Model: Getinge Steam Sterilizer Door
Affected:HS 5510 AC-1, HS 6613 DR-2, HS 5510 EC-1, HS 6610 ER-1, HS 6613T AR-1, HS 6613T AR-2, HS 6613 ER-2, HS 5510 ER-1
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Sterilization AB
Issue:The manufacturer received a report of a user injury involving the autoclave door. The sterilization technician activated the door lock while a basket was obstructing the door's movement. Rather than reopening the door and repositioning the basket, the technician attempted to manually adjust the basket while the door was closing. When the basket became dislodged, the door closed and resulted in a serious crushing injury to the technician's hand.
During further investigation, a second complaint involving a similar device was also identified.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/8/2026
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /