Detail of Recall

Type of Product:Device
Medsafe Reference:37166
Brand Name:Guldmann GH1 255kg SWLCeiling hoist
Model:GH1 F lifting module
Affected:units produced before 20 November 2020
Software version:
Recalling Organisation:Active Rehab Equipment
25 Rennie Drive
Airport Oaks
Auckland
Contact Information:0800 336 339
Manufacturer:V Guldmann A/S
Issue:Following a recent incident, it was identified that a GH1 F lifting module, which did not have the required software update and had an incorrectly installed Click Lock, fell onto a patient.
This may occur when the deinstallation is initiated and the installation cord is loosened without the GH1 F lifting module being released from the Click lock. In this case, the GH1 F lifting module could fall by the length of the loosened cord and in rare cases hit a patient or staff.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/8/2026