Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37166 |
| Brand Name: | Guldmann GH1 255kg SWLCeiling hoist |
| Model: | GH1 F lifting module |
| Affected: | units produced before 20 November 2020 |
| Software version: | |
| Recalling Organisation: | Active Rehab Equipment 25 Rennie Drive Airport Oaks Auckland |
| Contact Information: | 0800 336 339 |
| Manufacturer: | V Guldmann A/S |
| Issue: | Following a recent incident, it was identified that a GH1 F lifting module, which did not have the required software update and had an incorrectly installed Click Lock, fell onto a patient. This may occur when the deinstallation is initiated and the installation cord is loosened without the GH1 F lifting module being released from the Click lock. In this case, the GH1 F lifting module could fall by the length of the loosened cord and in rare cases hit a patient or staff. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/8/2026 |