Detail of Recall

Type of Product:Device
Medsafe Reference:37146
Brand Name:Drying Cabinet S-363
Model:Drying cabinet Double door and Drying cabinet Single door
Affected:S-363D200, S-363D400, S-363E300, S-363E400
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Disinfection AB
Issue:The manufacturer has identified a potential problem whereby hinge screws may loosen over time, causing the door of the drying cabinet to detach and fall. While testing equivalent to one year of use showed no signs of loosening, the exact timeframe for occurrence cannot be determined. A falling door may result in user injury; however, there is no identified risk to patients.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/7/2026