Detail of Recall

Type of Product:Device
Medsafe Reference:36883
Brand Name:Endo Therapy Disposable Sterile - Single Use Distal Cover
Model:Olympus Single Use Distal Cover - Model MAJ-2315
Affected:Multiple affected lot numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer conducted an investigation into the Single Use Distal Covers after receiving complaints of the distal cover unexpectedly detaching from the duodenoscope during use. In the past 12 months, the manufacturer has received complaints pertaining to the MAJ-2315 distal cover falling off within or outside patients. This included events considered to be Serious Injury, which involved foreign body retrieval and/or additional treatment. To address this issue, the manufacturer has developed a newly designed version of the Distal Cover (MAJ-2315), which contains a new attachment tool to help mitigate the risk of incorrect attachment of the distal cover to the scope. All older versions of this product should be returned and replaced with the newly designed version.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2026