Detail of Recall

Type of Product:Device
Medsafe Reference:36882
Brand Name:Mindray BeneHeart C/D series AED - non rechargable
Model:Mindray BeneHeart D1 AED
Affected:Serial numbers - FQ-55068195, FQ-55068196, FQ-58068556, FQ-59068749, FQ-59068750
Software version:
Recalling Organisation:Connected Healthcare Systems
6C Jack Conway Ave
Manukau
Auckland 2104
Contact Information:
Manufacturer:Shenzhen Mindray Biomedical Electronics Co Ltd
Issue:To help users better understand the procedure for switching patient types on the Mindray BeneHeart D1 Automated External
Defibrillator (AED), the manufacturer has updated the Operator's Manual with refined guidance.
When the device is set to paediatric mode and used for adult patients, there is a risk of ineffective defibrillation.
This Operator's Manual revision only adds supplementary information. No changes have been made to the product itself or its
operational procedures.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2026