Detail of Recall

Type of Product:Device
Medsafe Reference:36881
Brand Name:AirSeal® iFS Access Port with Trocar
Model:
Affected:Catalogue number - iASB5-150, iASB12-120, iASB12-100, iAS8-120LP, iAS8-100LP, iAS5-75LP, iAS5-120LP, iAS5-100LP, iAS12-150LPI, iAS12-120LPI, iAS12-100LPI
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:ConMed Corporation
Issue:The manufacturer has received an increase in reports of overpressure alerts with AirSeal Access Ports, which in some cases have resulted in delay of treatment. During a review of complaint data, the manufacturer identified that a previous manufacturing process update may, under certain conditions, contribute to overpressure alerts. In response, the Instructions for Use (IFU) have been clarified to ensure proper device tightening.
In the event of an overpressure condition, AirSeal® is designed to respond automatically via a built-in venting system, which activates when the pressure exceeds the set point by more than 3 mm Hg for more than 3 seconds. When there is improper device tightening, the overpressure alert may be erroneous, representing a higher-than-expected pressure within the system between the AirSeal® iFS and the AirSeal® access port but not a higher-than-expected pressure within the patient.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2026