Detail of Recall

Type of Product:Device
Medsafe Reference:36880
Brand Name:Atellica CH Enzymatic Hemoglobin A1c (A1c_E)
Model:
Affected:All in-date lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed, through internal investigation of customer complaints, the potential for intermittent well-to-well bias affecting the A1c_E/A1c_H assay when used on the Atellica CH and Atellica CI systems. This observed bias does not affect every A1c_E reagent lot or reagent pack. However, affected wells cannot currently be identified prospectively, therefore all in-date lots are considered potentially impacted until further notice. Under certain circumstances, this problem may result in:
1. Falsely depressed patient results when lot calibration is performed using an affected well and patient testing is subsequently performed using an unaffected well.
2. Falsely elevated patient results when lot calibration is performed using an unaffected well and patient testing is subsequently performed using an affected well.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2026