Detail of Recall

Type of Product:Device
Medsafe Reference:36868
Brand Name:Incubator, infant, stationary
Model:Giraffe OmniBeds - All models, Giraffe OmniBed Carestations - model 2082844-001-XXX, Field Replaceable Unit (FRU) - model 6600-1056-400 and 6600-1461-500
Affected:Canopy Soffits and fasteners on Giraffe OmniBed and Giraffe OmniBed Carestation Devices
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2026