Detail of Recall

Type of Product:Device
Medsafe Reference:36832
Brand Name:Automated Impella controller with Impella Connect
Model:Impella Controller, Impella Optical Controller, Optical, AIC, Impella Connect, Optical AIC w/Impella Connect, Optical, AIC, Impella Connect
Affected:Affected serial numbers - AIC with software version 6.0.1–6.0.3 and =7.1
Software version:
Recalling Organisation:Bio-Excel (NZ) Limited trading as Bio-Excel (Australia) Pty Ltd
Unit C6, 9 Tait Place
Albany
Auckland 0632
Contact Information:
Manufacturer:Abiomed Inc
Issue:The manufacturer has identified a software problem affecting the Automated Impella Controller (AIC) when used with a left
ventricular Impella device. The problem may occur in patients with prolonged minimal or absent pulsatility, defined as more than 80
minutes with an aortic placement signal below 12 mmHg.
If a sudden change in left ventricular pressure occurs while left ventricular pressure calculation is active, typically at support
levels above P-3, the AIC may restart due to an internal error. Disabling the aortic placement signal and pressure display does
not prevent this.
During a restart, the screen becomes blank without warning and the pump stops, leaving the patient unsupported and at risk of
cannula regurgitation. After restarting, the pump automatically returns to the previous performance level. Recovery to the prior
support level may take up to 75 seconds based on preliminary data. Replacing the AIC does not resolve this problem.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/5/2026