Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36813
Brand Name:Aidacare Aspire Mobile Patient Lifter - Powered
Model:Aidacare Aspire A205 Aluminium Patient Lifter
Affected:Item Numbers: LSS390610
Software version:
Recalling Organisation:Compliance Management Solutions NZ
13 Fernbank Lane
Greenhithe
North Shore City
Auckland 0632
Contact Information:
Manufacturer:Aidacare Pty Ltd
Issue:The manufacturer has identified that improper attachment of slings to the lifter's sling bar can cause patients to fall, sustaining severe injuries or death. Aidacare has enhanced the instructions for use (IFU) to strengthen existing instructions around correct and incorrect sling attachment methods and pre-use checks.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2026