Detail of Recall

Type of Product:Device
Medsafe Reference:36779
Brand Name:Radio-frequency ablation system probe cannula, reusable
Model:TCNTM (Thermo Couple Nitinol)
Affected:All batches
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Boston Scientific Corporation
Issue:The manufacturer has determined that the Thermo Couple Nitinol (TCN) and Cosman Straight Kit (CSK) Reusable RF Electrodes may
not consistently meet the expected performance levels throughout the specified 20 sterilisation cycles for re-use. This determination was based on recent review of internal laboratory testing. Complaint data indicates that the most frequently reported problems relate to electrode not heating, temperature-reading problems and device damage, which have led to prolonged
procedure. A small number of cases (0.15%) reported unintentional burns, allergic reactions, increased pain, or the need for additional intervention to retrieve potentially unrecovered parts (0.03%)
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/5/2026