Detail of Recall

Type of Product:Device
Medsafe Reference:36740
Brand Name:Olympus - Ligating device, single-use (Japan)
Model:SINGLE USE LIGATING DEVICE POLYLOOP - Model - HX-400U-30/ N5382130
Affected:Lots - 35V to 39V, 3XV, 3YV, 3ZV, 41V to 49V, 4XV, 4YV, 4ZV, 51V to 59V
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:In a communication dated 4 November 2025 (Medsafe Ref# 36102), the manufacturer informed users of complaints that the ligation loop could fail to release as intended, becoming unintentionally anchored around patient anatomy. Globally, serious injuries have been reported, with no deaths. The manufacturer determined the issue was caused by inadvertent or intentional movement of the yellow tube joint away from the device handle during use, particularly when moved to prematurely tighten the loop, which can prevent proper loop release.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/4/2026