| Type of Product: | Device |
| Medsafe Reference: | 36640 |
| Brand Name: | Aidacare Aspire Mobile Patient Lifter - Powered |
| Model: | Aidacare Aspire Aura and Nova Lifters |
| Affected: | Item Numbers: LSS390615 & LSS390652 - (serial numbers prior to XXX-PO01260162-0000) |
| Software version: | |
| Recalling Organisation: | Compliance Management Solutions NZ 13 Fernbank Lane Greenhithe North Shore City Auckland 0632 |
| Contact Information: | |
| Manufacturer: | Aidacare Pty Ltd |
| Issue: | The manufacturer has identified that the bolts securing the leg-spread actuators to the lifter frame can gradually loosen over time with regular use. If not addressed, this may lead to bolt detachment. As a result, one or both legs may become decoupled from the actuators and move freely when push or pull forces are applied to manoeuvre the lifter. If this occurs, users may experience difficulties when moving the lifter - whether or not a patient is suspended. The lifter may veer or pull noticeably toward the affected leg when force is applied, impacting safe and controlled operation. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/3/2026 |