Detail of Recall

Type of Product:Device
Medsafe Reference:36547
Brand Name:Azurion
Model:Philips Azurion R2.1.10 and 2.2.10 Systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Angiographic X-Ray
Model:Philips Azurion R2.1.10 and 2.2.10 Systems.
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified six software problems affecting Azurion R2.1.10 and R.2.10 systems that may result in loss of imaging (Xray) functionality, and/or loss of motorised movement, and/or incorrect image content, and/or loss of data.
Problem 1- System keeps restarting
Problem 2- AMC triple drive firmware – leading to movement problems
Problem 3- C-Partition of Suite PC running out of free space
Problem 4- System remains in continuous restart mode after the start-up
Problem 5- Misalignment of Marker Tool Overlay
Problem 6 - 1- Longitudinal Position Error – Applicable only to
Azurion systems with Poly-G3 frontal stand.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/2/2026