Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36310 |
| Brand Name: | Anaesthesia Work Stations |
| Model: | Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | There is the potential for unexpected shutdown of Carestation 600 and 700 Series Anaesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anaesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings. If this situation is not identified and addressed by the user, the loss of ventilation may be life threatening. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/12/2025 |