Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 35834 |
| Brand Name: | DxI 9000 Access Immunoassay Analyzer |
| Model: | |
| Affected: | SW 1.22 and below |
| Software version: | |
| Recalling Organisation: | Beckman Coulter Australia Pty Ltd 1 Dale Street Brookvale NSW 2100 Australia |
| Contact Information: | |
| Manufacturer: | Beckman Coulter Inc |
| Brand Name: | DxC 500i Clinical Analyzer |
| Model: | Access 2 Module, DxC 500i - |
| Affected: | SW v1.3.5 and lower |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has determined that the analyzer may fail to apply a dilution factor to tests associated with a manually diluted sample when the tests are ordered through the Test Order Entry page and a unique sample ID is not used. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/9/2025 |