Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:35667
Brand Name:Ingenia, Achieva Series
Model:
Affected:Ingenia 1.5T 781341 Ingenia 3.0T 781342 Ingenia Ambition S 781359 782108 Ingenia Elition S 781357 782106 Ingenia Elition X 781358 782153
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:PMS Nederland MR Systems
Model:Philips MR Systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified an issue where component failures in the Gradient Coil of the affected MR systems listed in Section 3 may act as a heat source with a potential to produce smoke and/or fire.
If a component failure occurs, the user may observe the following:
Repeated scan abort error messages due to gradient amplifier fault detection, described in Section 4.
Unusual noise from the system in the examination or technical room
Burning smell
Smoke and/or fire
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/8/2025