Detail of Recall

Type of Product:Device
Medsafe Reference:35627
Brand Name:DxC 500i Clinical Analyzer
Model:DxC 500 AU Module w/ISE, DxC 500i and DxC 500 AU Module, DxC 500i
Affected:Software - SW 1.3.3 and 1.3.4
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that on the DxC 500i Clinical Analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the Clinical Chemistry constituent tests, the sample status of the Clinical Chemistry test will remain In Progress with no error events.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/8/2025