Detail of Recall

Type of Product:Device
Medsafe Reference:35560
Brand Name:Angiographic X-Ray
Model:Philips Azurion Systems
Affected:Multiple models
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Azurion
Model:Philips Azurion R1.x and R2.x systems
Affected:Multiple model numbers
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified a software issue in the internal communication process between the system software and the X-ray generator firmware. This issue may be triggered under either of the following situations:
Situation 1 - Pedal Tap: When the foot switch is pressed and released rapidly without generating X-rays (due to a short duration of contact), the system may enter a state where X-ray generation is inhibited. When this occurs, the system does not show any message to the user indicating that X-ray is not possible.
Situation 2 – Phase Fault: Azurion systems are designed to automatically recover from phase faults (a phase fault is an abnormal condition in which one or more phase voltages drop to (near) zero). When a phase fault occurs, the X-ray generator software initiates an automatic recovery process to restore full system functionality, during which imaging (X-ray) functionality is not available for approximately 5 seconds.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/7/2025