Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 35496 |
| Brand Name: | MRI system, full-body, superconducting magnet |
| Model: | SIGNA Artist, SIGNA Voyager, SIGNA Artist EVO and SIGNA Hero |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Tianjin Company Ltd |
| Brand Name: | MRI system, full-body, superconducting magnet |
| Model: | Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | The manufacturer has identified that the planned maintenance procedures for verifying the functionality of the caster locks on MR patient tables may not have been consistently performed. This functional check is a proactive safety measure designed to detect potential caster lock failures. If not identified, such failures could lead to unintended table movement and unstable patient support when the table is not docked to the system. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/7/2025 |