Detail of Recall

Type of Product:Device
Medsafe Reference:35429
Brand Name:CRHF - CareLink SmartSync Device Manager System
Model:Multiple Software Components
Affected:Update SmartSync (to version 4.2.3 or higher
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:A software update for Medtronic CareLink™ SmartSync™ Device Manager (SmartSync) is now available. The SmartSync update aligns the Abort button behavior within the EP Study screen with legacy Medtronic programmers. Once the update is applied, the Abort button cancels any therapy in progress initiated via the EP Study screen. In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory