Detail of Recall

Type of Product:Device
Medsafe Reference:35329
Brand Name:Ingenia, Achieva Series
Model:Philips MR systems with SW versions R11.1 and R12.1
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Ingenia, Ingenia Elition, Ingenia Ambition, MR 7700 MRI Systems
Model:MR systems operating on software versions R11.1 and R12.1
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:26/05/2026: Update - The manufacturer has identified potential alignment errors in the cross-reference line functionality when reviewing images generated with the MobiView application.

The manufacturer has identified potential alignment errors in the cross-reference line functionality when reviewing images generated with the MobiView application.
A. The cross-reference line and/or box may be shown at an incorrect location when displayed on a MobiView generated image.
The alignment errors may occur only if both of the following conditions are met:
Multi-station scans that require stitching 3 or more station images, and
Scan protocol is used where stations have different voxel size or field of view (i.e. voxel size or field of view is customized based on anatomy per station).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/5/2026