Type of Product:Device
Medsafe Reference:35238
Brand Name:Soltive Fibres
Model:Models - TFL-FBX150BS 150, TFL-FBX150BS 200, TFL-FBX150BS 365, TFL-FBX150BS 550, TFL-FBX150BS 940, Soltive Pro Superpulsed Laser Unit, Soltive Premium Superpulsed Laser Unit
Affected:all lot numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Gyrus ACMI Inc
Issue:The manufacturer has identified reportable malfunctions and serious injuries, including breakage of a fibre that resulted in a burn to the hand of the user and a single-use fibre that caught on fire causing a procedural delay. There is no evidence that material within the operating room progressed to a fire in any of the reported incidents.
The preliminary results of the investigation indicate that a fibre break, at the strain relief or along the length of the fibre, is the cause of, or contributes to, these incidents. The manufacturer is continuing to investigate the issue and will take appropriate actions regarding the investigation findings.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Wholesale
Date Commenced:19/5/2025