Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:34782
Brand Name:DxC 500i Clinical Analyzer
Model:With ISE - DxC 500 AU Module w/ISE, DxC 500i, Access 2 Module, DxC 500i. Without ISE - DxC 500 AU Module, DxC 500i, Access 2 Module, DxC 500i
Affected:SW 1.3.0 and 1.3.2
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that the DxC 500i Clinical Analyzer might assign a test result to an incorrect sample ID under the following workflow conditions. After loading a barcoded sample tube (Sample A) onto the DxC 500i Clinical Analyzer, all tests ordered for that barcoded sample tube are locked into the specific position of that barcoded tube on the analyzer (for example, the position on the DxLab Rack, Carousel Rack or position in the STAT table) until all tests are completed or cancelled for that sample tube. A problem might occur if, for any reason, all tests ordered for that barcoded sample tube (Sample A) are not completed or remain “in pending”, and the operator attempts to analyze a second sample tube (Sample B) in that same position on the rack or STAT table.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/3/2025