Detail of Recall

Type of Product:Device
Medsafe Reference:34749
Brand Name:Anaesthesia Work Stations
Model:Aisys CS² and Carestation 750/750c anesthesia devices with affected electronic gas mixer assemblies.
Affected:Aisys CS² and Carestation 750/750c anesthesia devices with electronic gas mixer assemblies manufactured between March 31, 2023, to June 24, 2024.
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware of a potential problem in certain anesthesia devices manufactured between March 31, 2023, to June 24, 2024, which contain an electronic gas mixer assembly.

The potential problem does not affect gas delivery or gas flow rate to the patient and does not impact inspired oxygen concentration.

Over time potentially impacted gas mixers can develop an internal fault, resulting in a hissing noise. The hissing noise is audible upon addition of a balance gas (air or nitrous oxide (N2O)).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/2/2025