Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 34747 |
| Brand Name: | DxC 500i Clinical Analyzer |
| Model: | DxC 500 AU Module w/ISE, DxC 500i, Access 2 Module, DxC 500i, DxC 500 AU Module, DxC 500i, Access 2 Module, DxC 500i |
| Affected: | SW 1.3.0 and 1.3.2 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has determined that the system cannot recalculate a calculated result when results are rerun within the calculated test. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/2/2025 |