| Type of Product: | Device |
| Medsafe Reference: | 34596 |
| Brand Name: | Liquichek Unassayed Chemistry Control |
| Model: | |
| Affected: | Level 1 - lot - 92971, 92981. Level 2 - 92972, 92982. MultiPak - lot - 92970, 92980 |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Bio-Rad Laboratories Inc |
| Issue: | Thawed Opened vial storage stability claim for Total Acid Phosphatase in Liquichek Unassayed Chemistry Control has been revised to 6-days. All other claims for these lots of Liquichek Unassayed Chemistry Control remain unchanged. The STORAGE AND STABILITY section has been revised to indicate the new stability claims for Total Acid Phosphatase. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/1/2025 |