Detail of Recall

Type of Product:Device
Medsafe Reference:34168
Brand Name:Angiographic X-Ray
Model:Philips Allura Xper R7.6- R8.1 systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:22/10/25: Update - The maufacturer initiated Corrective Action 2024-IGT-BST-010 to correct Allura R7.6 and R8.1 systems that have exceeded their 10-year service life, and for which the propeller motor clamping bolt and the X-ray tube locking bolt had not been replaced. This action was taken due to the potential risk of bolt failure. However, following further investigation, the manufacturer has confirmed that the initially suspected failure mechanisms—hydrogen embrittlement and fatigue - do not apply. Therefore, replacement of these bolts is not required.

The manufacturer has identified potential safety issues with delayed LTE kit installations for Allura R7.6-R8.1 systems exceeding 10 years of life. The kit includes a propeller motor clamping bolt and the X-ray tube locking bolt. If one of these bolts breaks, part of the system will not function as expected.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/10/2025