Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 34087 |
| Brand Name: | HD and PD disposables 35000 IIa Zurich |
| Model: | Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp), Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp – Extra Short), Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp – Extra Long) |
| Affected: | Product code - R5C4482, R5C4483, R5C4484 |
| Software version: | |
| Recalling Organisation: | Vantive Limited 7-9 Niall Burgess Road Mt Wellington Auckland 1060 |
| Contact Information: | |
| Manufacturer: | Baxter Healthcare SA (Switzerland) |
| Issue: | 9/19/26: Update - The manufacturer has provided additional guidance and recommends continuing to use peroxide-cured silicone tubing sets. For patients under 6 months of age (approximately 3.5 kg), shorter transfer sets should be prioritised where platinum-cured silicone tubing or other alternatives are unavailable. Issue 1: An increase in complaints has been reported involving separation of the female connector from the transfer set during connection or disconnection. The sterile fluid path remains intact, with no resulting risk of microbial contamination. Issue 2: The MiniCap Extended Life PD Transfer Set contains peroxide-cured silicone tubing. The manufacturer is assessing a potential patient safety risk and is providing interim information while the evaluation is ongoing. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/10/2026 |