Detail of Recall

Type of Product:Device
Medsafe Reference:34087
Brand Name:HD and PD disposables 35000 IIa Zurich
Model:Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp), Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp – Extra Short), Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp – Extra Long)
Affected:Product code - R5C4482, R5C4483, R5C4484
Software version:
Recalling Organisation:Vantive Limited
7-9 Niall Burgess Road
Mt Wellington
Auckland 1060
Contact Information:
Manufacturer:Baxter Healthcare SA (Switzerland)
Issue:9/19/26: Update - The manufacturer has provided additional guidance and recommends continuing to use peroxide-cured silicone tubing sets. For patients under 6 months of age (approximately 3.5 kg), shorter transfer sets should be prioritised where platinum-cured silicone tubing or other alternatives are unavailable.
Issue 1: An increase in complaints has been reported involving separation of the female connector from the transfer set during connection or disconnection. The sterile fluid path remains intact, with no resulting risk of microbial contamination. Issue
2: The MiniCap Extended Life PD Transfer Set contains peroxide-cured silicone tubing. The manufacturer is assessing a potential patient safety risk and is providing interim information while the evaluation is ongoing.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/10/2026